Digiquartz Data Storage (DDS)

 

Digiquartz Data Storage (DDS) is an onboard serial EEPROM chip that contains calibration and transducer information. This information can be accessed electronically to provide plug and play pressure transducer interchangeability.

Transducers with DDS can be identified by the presence of additional wires to a standard transducer configuration. The sole purpose of DDS is to store transducer configuration information. Use of DDS is optional and does not affect pressure measurement and accuracy.

Series 4000 Sensor with DDS wires (side view)

Series 4000 Sensor with DDS wires (top view)


DDS wires have the following color codes.
 
Wire Color Signal
   White/brown    Serial Clock (SCL)
   White/Violet    Serial Data (SDA)
   Black    Ground

The maximum recommended length of the SDA and SCL wires is 18 inches.  Serial communication with the serial EEPROM may become unreliable at longer wire lengths. 

The information stored in DDS memory is organized as in the DDS Memory Map below.  An additional 4K bytes memory has been allocated for customer application use.

Please review the transducer installation manual about programming details and power requirements.

Warning: Voltages outside the range of -0.3 to 4.3 volts must not be applied to the SDA and SCL signal lines.  Doing so will damage the DDS serial EEPROM.

 

Digiquartz® Data Storage (DDS) Memory Map

 