|
Calibration
of MET3A Temperature Measurement System |
|
|
|
Summary |
|
|
The
temperature measurement system of three MET3A prototype units
were calibrated over the range of 50 to +60° C.
Subsequent testing demonstrated system temperature
conformance that was well within +/- 0.1° C, referenced to a
calibrated and NIST-traceable reference thermistor probe.
|
| |
|
Background |
|
The MET3A
features a precision thin-film RTD to sense temperature.
This technology requires a calibration technique that is
somewhat different than the technique commonly used to calibrate
temperature probes. For
MET3A temperature calibration, a thermal chamber is used to
control environmental temperature, and an independently
calibrated and NIST traceable thermistor-based thermometer is
used as a temperature reference.
The MET3A
humidity/temperature sensor is surrounded by a small thermal
mass that features a bore designed to accept a standard 0.125
diameter temperature reference probe.
The bore terminates near the humidity/temperature sensor,
assuring isothermal contact between the sensor and the
temperature reference probe.
The MET3A RTD sensor and electronics can therefore be
simultaneously calibrated by taking reference temperature and
MET3A temperature measurements at various stable temperatures
over the range of 50 to +60 °C, and adjusting the MET3A
temperature coefficients to correct for combined temperature
sensor and electronic nonlinearities.
This process yields system temperature measurement
accuracy of +/- 0.1 degrees C or better over the temperature
range of 50 to +60 degrees C.
|
|
Data
Acquisition |
|
The Device Under Test (DUT)
is placed in the thermal chamber, and is connected via its
serial interface to a PC running Digiquartz Interactive software
(DQI). The unit is
queried to ensure that it is configured with default calibration
coefficients. The
reference thermistor probe is inserted into the thermal mass,
and is read by a Hart Scientific Model 1560 Thermometer Readout.
The traceable accuracy of the thermistor probe and
thermometer readout is +/- 0.016 degrees C over the range of
50 to +60 degrees C, based on a multi-point calibration
performed at the Hart Scientific metrology lab.
Calibration
data are acquired at 50, -25, 0, 25, and 60° C.
At each temperature, the indicated temperature from the
DUT and the reference thermistor are monitored until thermal
equilibrium was reached.
Thermal equilibrium is typically
reached after soaking the DUT at the desired temperature for two
hours. Once the DUT
temperature has stabilized, several DUT and reference thermistor
temperature measurements are taken, averaged, and recorded.
|
|
Data
Analysis |
|
At each
temperature, the equation T = E + Fx + Gx2 is solved
for x, where T is the indicated MET3A temperature, and E, F, and
G are the default MET3A coefficients.
The resulting values of x are fitted to the corresponding
reference thermistor temperature measurements using a quadratic
model, yielding a new set of corrected coefficients.
|
|
Temperature
Conformance Test |
|
The DUT is
configured with the corrected
coefficients. Conformance data are acquired at
50, -25, 0, 25, and 60° C.
Data are typically acquired over three complete
temperature loops. At
each temperature, the indicated temperature from the DUT and the
reference thermistor are monitored until thermal equilibrium is
reached. Thermal
equilibrium is typically reached after soaking the DUT at the
desired temperature for two hours.
Then, several DUT and reference thermistor temperature
measurements are taken and averaged, and the averages are
recorded. Temperature
error at each test temperature is calculated by subtracting the
reference thermistor temperature from the indicated DUT
temperature.
|
|
Calibration
Results |
|
Figures 13
show the results of temperature calibrations performed on three
MET3A prototypes. Conformance data were typically acquired over
three complete temperature loops.
In all three cases, system temperature conformance to a
highly accurate and traceable temperature reference probe are
well within +/- 0.1° C over the range of 50 to +60 °C.
Refer to Figures 13.
|
|
 |
|
Figure 1. SN 73354 |
|
 |
|
Figure 2. SN 74906 |
|
 |
|
Figure 3. SN 73382 |
|
©2004
Paroscientific, Inc.
|
|