|
The
MET3A is designed to provide highly accurate temperature
measurements using a fan-aspirated ventilation system.
Non-aspirated systems using radiation shields and louvered
screens are subject to large errors.
A
MET3A was instrumented with temperature sensors and measurements
were taken of the temperatures of the outer shield, inner shield
and the isothermal block. The MET3A was placed in the sun and
the temperatures were recorded. The solar radiation flux density
at the time of the test was approximately 850 W/m2.
The wind speed was fluctuating between 0.4 and 1.0 meter/sec.
The measured temperature rise of the outer shield was 3°C.
Tests
were performed on the MET3A to determine the effects of solar
heating on measurement accuracy. To accomplish this, a
side-by-side comparison was made of two MET3As. These devices
were mounted outdoors over a grassy area. Data was recorded
continuously throughout the test. Both systems were exposed to
the sun initially. After allowing time to confirm that they
track each other, one unit was shaded from the sun. After an
hour the shade was removed. The difference between the readings
while the one unit was shaded is the error due to solar
radiation.

The
wind during this test fluctuated between 0.5 and 1.0 m/s. The
solar radiation flux density ranged from 620 to 970 W/m2
during the test with an average of 820 W/m2.
The
following graphs show the effects of solar radiation equal to
0.2ºC.

Reviewing
specifications from the RM Young multi-plate radiation shield,
the errors in equivalent conditions (wind speed 0.5 to 1 m/s and
solar energy of 820 W/m2) are between 1 to 1.5 ºC. With wind of
less than 0.5 meter/sec, the errors increase dramatically. These
numbers are confirmed by the Andersson/Mattisson report,
Reference 1.
Paroscientific
has tested the multi-plate radiation shield and observed similar
results. A comparison with the MET3A shows daily overshoots that
can exceed 2 ºC during calm and sunny periods. It can be seen
on the following graph that the solar heating of the shields
cause the temperature reading to increase more rapidly than the
actual air temperature and to overshoot by over 2 ºC.

The
conclusion is the aspirated MET3A is 5 to 7 times more accurate
than the non-aspirated systems. |