STABILITY TEST REPORT OF 760-15A PORTABLE FIELD STANDARD


by 
Dr. Theo P. Schaad

Subject:  Long-term Stability of S/N 34260(PA4), Model 760-15A

This unit was originally calibrated here Feb 1989, tested for stability in 1990 by Det Norske Met Inst. (KNMI study), and was later used as demo equipment all over the world (including Erpatec Switzerland, Micromeritics Georgia, Ammo Eng. Israel, and Canagrad Surveys Canada).  I tested the unit on 8 July 1994 at nominal 14.5 psia for zero stability.  The unit was reconfigured to the original calibration coefficients. 

Time elapsed:  5.4 years

Measurement 1: 

The unit was tested with the CEC Dead Weight Tester, Piston/Cylinder, Weight Set 2.  This set was last calibrated 2 Feb 1993 with a traceable accuracy of 0.01% of reading, or +/-0.00145 psi at the tested pressure value. 

Applied pressure:           14.50449 psia

Indicated pressure:         14.50336 psia

Difference:                           -0.00113 psia

Relative diff:                       -0.0076% fs

The original calibration had been done with the CEC Dead Weight Tester, Piston/Cylinder 1, Weight Set 1.  The calibration was based on the original CEC values (1987?) with the masses remeasured in 1988.  The accuracy had been stated as 0.015% of reading, or +/- 0.00218 psi at the tested value. This accuracy is largely independent of the current value.  Realistically, the two dead weight tester error should be added in this measurement to +/-0.00363 psi.  The typical repeatability achieved on measuring a barometer is 20 ppm or +/-0.0003 psi.  Hence the total stability should be expressed as:

Stability in 5.4 years:  -0.00113 +/-0.0003 +/-0.00363 psi

where the first error expresses the precision of the measurement, and the second the known accuracy.  This gives a total range of between –0.0051 and +0.0028 psi. 

Relative slope:  between –0.0063% fs and +0.0034% fs per year.   

Measurement 2:  The unit was tested with the CEC Dead Weight Tester, Piston/Cylinder 1, Weight Set 1.  The old calibration coefficients and weight values (1989) were used. 

Applied pressure:            14.49936 psia

Indicated pressure:         14.49705 psia

Difference:                      -0.00231 psia

Relative diff:                    -0.0154% fs      

The success of this measurement depends how much the applied pressure of the dead weight tester has changed over the years.  Since we mostly used another piston set since 1990, I expect that the piston area of Piston 1 is essentially unchanged.  Again, as above, a repeatability error of 20 ppm should be considered. 

Stability in 5.4 years:  -0.00231 +/-0.0003 psi

Relative slope:  between –0.0032% fs and –0.0025% fs per year.